New LESAPLUS

Advion TriVersa LESA

Liquid Extraction Surface analysis coupled with LC (LESAPLUS LC)
LESAPLUS LC experiments can be executed through the Developers Kit software extension and allows for the added benefit of separation. Spatial resolution can be improved to 400 µm on target and once the extract is placed in the loop of the on-board 6 port valve, an nLC/MS experiment can commence.

 

Typical Applications

  • PK/PD analysis of small molecule drugs
  • Spatially resolved lipidomics
  • Peptidomics

Benefits

  • Direct Analysis from surface of interest
  • 400 µm spatial resolution
  • Liquid extraction and nESI based analysis
  • High sensitivity due to added chromatographic separation
  • Analysis of complex mixtures and isobaric compounds
  • Derivatization/Enzymatic digestion prior to LESAPLUS experiments

TriVersa NanoMate Success Stories

See more success stories »

University of Southern Denmark

"Essential for any Lipidomics Laboratory."

Christer S. Ejsing, PhD, Associate Professor

Read More »

University of Birmingham

"The TriVersa NanoMate revolutionized nanospray."

Dr. Helen Cooper, Senior Lecturer, School of BioSciences

Read More »

Max Planck Institute of Molecular Plant Physiology Golm, Germany

"I can’t imagine running these analyses without the TriVersa NanoMate"

Dr. Patrick Giavalisco, Research Group Leader

Read More »

School of Chemistry, University of Bristol – Bristol, UK

"Advion’s technology gives you what you want very quickly."

John Crosby, Senior Lecturer, Organic & Biological Chemistry

Read More »

Recent TriVersa NanoMate Publications

See more publications »

TriVersa NanoMate Application Notes

See more application notes »

Shotgun Lipidomics Analysis

Infusion based lipidomics analysis (shotgun Lipidomics) is an analysis strategy that involves the liquid extraction of lipids from samples such as blood, tissue homogenates or dried blood spots. This approach is suitable for a wide range of biomedical questions of lipid differentiation and lipid flux in the field of metabolic disease, neurological disorders, cancers or eye…

Read More »

Microorganism Surface Analysis Using the TriVersa NanoMate for Liquid Extraction Surface Analysis of Bacteria and Fungi

Liquid Extraction Surface Analysis Mass Spectrometry (LESA-MS) has been utilized for determination of small molecules, lipids and proteins from a variety of surfaces such as tissue sections, plant material, medical devices, TLC plates or DBS cards. It allows direct surface analysis without sample preparation from spot sizes of 1 mm diameter by aspiration of extraction…

Read More »

Pesticides from Food Surfaces Using LESA PLUS Liquid Extraction Surface Analysis PLUS LC Separation

Residues and contaminants in the food chain are an increasing problem due to high production volumes, large area distribution and import/export of food items across the world. The novel LESAPLUS surface analysis approach combines the standard liquid extraction surface analysis with an additional step of a nano liquid chromatography separation. This combination is ideally suited…

Read More »

Using the TriVersa NanoMate with LESA for Analyte Profiling in Material Science

Surface analysis techniques have gained significant attention over the last couple of years with LESA being an interesting alternative to techniques such as SIMS or MALDI imaging. This note demonstrates practical applications for LESA in the material sciences.

Read More »