Specifications

Sample Interface

Sample Introduction

  • Peristaltic Pump: 4 channels, 12 rollers
  • Spray Chamber: T-controlled, Cyclonic, Quartz
  • Spray Chamber T: -10 – 35 ºC
  • Nebulizer: PFA and Quartz
  • Nebulizer Flow Rate: 1 mL/min

Plasma Torch

  • Rf Generator Frequency: 27 MHz
  • Rf Generator Power: 500-1600 W

Optics

  • Cones: Ni and Pt tipped Cu
  • Sample Cone i.d.: 1 mm
  • Skimmer Cone i.d.: 0.4 mm
  • Extraction Cone i.d.: 1.02 mm
  • Lens: Einzel
  • Optics: 90º quad deflector

Mass Spectrometer

Mass Spectrometer

  • Mass Range: 2-260 u (212 u for SOLATIONE)
  • Resolution: 0.7-0.9
  • Scan Speed: 5000 u/s
  • Abundance Sensitivity m-1: 2.5e-6
  • Abundance Sensitivity m+1: 2e-7
  • Mass Stability: <0.05 u / day

Collision Cell

  • Collision Cell Type: Octupole
  • Kinectic Energy Discrimination: Yes
  • No. Gases: 1 (He)

Detector

  • Dwell Time: 100 µs
  • Linear Dynamic Range: > 9 orders

Performance Specifications

Sensitivity

  • 7Li: >20 Kcs/ppt
  • 9Be: >3 Kcs/ppt
  • 115In: >50 Kcs/ppt
  • 205Ti: >40 Kcs/ppt
  • 238U*: >40 Kcs/ppt

*Available for SOLATIONHM models only

Detection limits:

  • 7Li: <1 ppt
  • 115In: <0.5 ppt
  • 209Bi: <0.5 ppt

Other Specifications:

  • CeO/Ce: <2.5%
  • Ba++/Ba+: <3%
  • Background cps (@ m/z 212): <4%
  • Stability, 20 min: <3%
  • Stability, 2 hrs: <4%
  • 107Ag/109Ag: <0.5%

Dimensions and Power Requirements

  • Size: 90 x 66 x 70 cm (W x D x H)
  • Volume: 0.42 m3
  • Weight: 115 Kg (250 lbs.)
  • Argon : 24 L/min (99.95%)
  • Cell Gas: 10 mL/min (99.999% He)
  • Exhaust (150mm): 5-7 m3/min
  • Power: 200-240 V, 30 A
  • Operating Temperature: 15-30 ºC (59-86 ºF), performance specifications require ±2 ºC from calibration temperature
  • Operating Humidity: 10% and 80% RH, con-condensing